THE BIURET METHOD FOR THE DETERMINATION OF TOTAL PROTEIN USING AN

Download Introduction. One commonly used method for determining the total protein in a sample is the Biuret method. The Biuret method is based on th...

0 downloads 486 Views 390KB Size
Application Note: 51859

The Biuret Method for the Determination of Total Protein Using an Evolution Array 8-Position Cell Changer Nicole Krueziger Keppy, Michael W. Allen, Ph.D, Thermo Fisher Scientific, Madison, WI, USA

Introduction Key Words • Biuret Method • 8-Position Cell Changer • Total Protein • UV-Visible Spectroscopy

One commonly used method for determining the total protein in a sample is the Biuret method. The Biuret method is based on the complexation of Cu2+ to functional groups in the protein’s peptide bonds as shown in Figure 1. The formation of a Cu2+-protein complex requires two peptide bonds and produces a violet-colored chelate product which is measured by absorption spectroscopy at 540 nm. Over a given NH 2 NH 2 concentration range, the measured absorption at 540 nm is linear with respect to the NH NH concentration of total protein. This relationship Cu 2+ allows a standard curve to be created that is NH 2 NH 2 used to calculate the Figure 1: Biuret reagent reacts with an concentration of an alkaline solution of CuSO4 to form a violet unknown sample. chelate compound

Experiment and Results The Biuret reagent was prepared by adding 3 g of CuSO4• 5H2O and 9 g of sodium potassium citrate to 500 mL of 0.2 N NaOH solution, followed by the addition of 5 g of KI. The resulting solution was then brought to a total volume of 1 L with 0.2 N NaOH. Alternatively, the Biuret reagent is available from a variety of sources including Thermo Fisher Scientific. Protein standards and the sample were prepared with saline solution (8.5 g/L) according to Table 1. 3.0 mL of Biuret reagent was added to each standard and sample, the solution was mixed well and incubated at room temperature for 30 minutes.

The standards and sample were analyzed using the Biuret method included in Thermo Scientific VISIONcollect software with biological tests. To begin the measurement, select the Biuret Method from the provided method files in Quantification Mode. The experimental method and 8-position cell changer set-up are shown in Figures 2 and 3 respectively.

Figure 2: Experimental Method Set-up using VISIONcollect

Test Tube Number Reagent

3.0 mg/mL BSA (mL) Protein Sample (mL) Saline Solution (mL) Final Concentration (µg/mL)

1

2

3

4

5

6

7

8



0.2

0.4

0.7

1.0

2.0

3.0

















1.0

3.0

2.8

2.6

2.3

2.0

1.0



2.0

0

200

400

700

1000 2000 3000

Table 1: Preparation of Protein Standards and the Sample

– Figure 3: Multi-Cell Method Set-up using VISIONcollect software

Standards 2 to 7 were measured at 540 nm using standard 1 as the reference sample, or blank. A linear fit was applied to the standard results in Table 2 to obtain the standard curve shown in Figure 4. The resulting calibration curve exhibits a linear relationship with a correlation coefficient (R2) of 0.9996. The unknown sample was measured in Quantification mode. Using the calibration curve the concentration of protein in the sample was calculated to be 1553 µg/mL, as shown in Table 2.

Solution

Standard 2 Standard 3 Standard 4 Standard 5 Standard 6 Standard 7 Unknown Sample

Concentration (µg/mL)

Absorbance

200 400 700 1000 2000 3000 1553

0.0238 0.0541 0.0862 0.1304 0.2514 0.3817 0.1972

In addition to these offices, Thermo Fisher Scientific maintains a network of representative organizations throughout the world.

Table 2: Results of Protein Standards and the Sample using Biuret Method

Conclusion Automated quantitative analysis of protein is performed quickly and easily using the Thermo Scientific Evolution Array UV-Visible spectrophotometer. The VISIONcollect™ software includes a pre-configured method for the Biuret assay, allowing further customization to individual laboratory protocols. Integration of the sample measurement and data analysis into VISIONcollect software saves time and improves laboratory throughput by eliminating post-measurement data manipulation. The 8-position cell changer enables measurements to be taken without exchanging the cells between measurements further increasing the efficiency of your methods. Figure 4: Biuret-Protein Complex Spectrum and Calibration Curve

Africa-Other +27 11 570 1840 Australia +61 2 8844 9500 Austria +43 1 333 50 34 0 Belgium +32 2 482 30 30 Canada +1 800 530 8447 China +86 10 8419 3588 Denmark +45 70 23 62 60 Europe-Other +43 1 333 50 34 0 Finland / Norway / Sweden +46 8 556 468 00 France +33 1 60 92 48 00 Germany +49 6103 408 1014 India +91 22 6742 9434 Italy +39 02 950 591 Japan +81 45 453 9100 Latin America +1 608 276 5659 Middle East +43 1 333 50 34 0 Netherlands +31 76 579 55 55 South Africa +27 11 570 1840 Spain +34 914 845 965 Switzerland +41 61 716 77 00 UK +44 1442 233555 USA +1 800 532 4752

www.thermo.com

©2009 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details.

Thermo Electron Scientific Instruments LLC, Madison, WI USA is ISO Certified. AN51859_E 12/09M

Part of Thermo Fisher Scientific