[email protected]
Minimum Requirements for Digital Radiography Equipment and Measurement Procedures by Different Industries and Standard Organizations Uwe Ewert and Uwe Zscherpel BAM Federal Institute for Materials Research and Testing, Berlin, Germany
8.3 Radiological Methods
ECNDT Praha, Oct. 2014
Requirements for RT-D Ewert and Zscherpel
1
Introduction - Film Radiography has been developed for quality assurance in industry over decades. - It is basis for contractual agreements of Supplier and Purchaser. - Digital Radiography shall achieve the same image quality or better, but with higher efficiency. - Selected image quality requirements of different standards as ASTM, ASME and ISO are compared. - The essential parameters for optimization and prediction of IQI visibility are discussed. - Requirements for equipment selection are derived for CR and DDAs. - Newly developed classification procedures for RT-D equipment are introduced.
8.3 Radiological Methods
ECNDT Praha, Oct. 2014
Requirements for RT-D Ewert and Zscherpel
2
Basic Requirements in Radiography Film
Requirement Standards
Exposure
Detector requirements
Maximum image unsharpness Image quality
8.3 Radiological Methods
Digital
ISO, CEN, ASTM, ASME
ISO, CEN
Exceed minimum optical density
Exceed minimum required SNRN (Normalized Signal to Noise Ratio)
Use detectors which achieve the required Use required film SNRN and system class or do not exceed the better detector unsharpness limit Do not exceed the Do not exceed the geometrical image unsharpness limits unsharpness limits Achieve IQI contrast Achieve required IQI sensitivity + duplex contrast sensitivity wire resolution Requirements for RT-D
ECNDT Praha, Oct. 2014
Ewert and Zscherpel
ASTM, ASME, MAI
DDAs: Exceed minimum required CNR CR: expose to EPS plateau (MAI) Do not exceed image unsharpness limits Do not exceed the image unsharpness limits Achieve required IQI contrast sensitivity 3
Key Technologies for Film Replacement • Computed Radiography (CR) with storage phosphor imaging plates • Digital Detector Arrays (DDA)
CR
8.3 Radiological Methods
ECNDT Praha, Oct. 2014
DDA
Requirements for RT-D Ewert and Zscherpel
4
Contrast Signal (base material)
Intensity
Intensity
Image Quality in Radiography - Influence of SNR and CNR -
Contrast Signal (base material)
Length
Length
Notch visible!
Notch not visible!
Contrast/Noise is high Signal/Noise is high
Notch is visible Contrast/Noise is low in the profile if Signal/Noise is low C > 2.5 Noise
8.3 Radiological Methods
ECNDT Praha, Oct. 2014
Requirements for RT-D Ewert and Zscherpel
5
Noise Sources in Radiographic Images Typical noise sources in digital radiography: 1. EXPOSURE CONDITIONS: Photon noise, depending on exposure dose (e.g. mAs or GBqmin). This is the main factor! SNR increases with higher exposure dose. 2. Limitation for the maximum achievable SNR: 1. DETECTOR: Structural noise of DDAs and Imaging Plates also called fixed pattern noise (due to variations in pixel to pixel response and inhomogeneities in the phosphor layer). 2. OBJECT: 1. Crystalline structure of material (e.g. nickel based steel, mottling) 2. Surface roughness of test object
8.3 Radiological Methods
ECNDT Praha, Oct. 2014
Requirements for RT-D Ewert and Zscherpel
6
Measurement of Contrast to Noise Ratio in DDA Practice by ASTM E 2698 Noise
I=contrast
• CNR shall be measured in the 4T hole for proof of image quality. • A minimum CNR of 2.5 is required by ASTM E 2698. • This value needs to be revised!
ASTM E 1025
I
8.3 Radiological Methods
ECNDT Praha, Oct. 2014
Requirements for RT-D Ewert and Zscherpel
CNR = 6.7 C = I = 473 Noise = 71 SNR = 155 7
Human Observer Model: Perception Threshold PT ASTM E 2698
50 µm pixel size Diameter 0.12 0.25 0.5
1
Noise = 1000 Signal = 30000
2 mm
requires CNR > 2.5 Needs revision!
Large area flaws are better visible than small ones at d - diameter same Contrast-to-Noise of hole Ratio
CNR = 2.5
CNR = 1.25
CNR = 0.625
Each column has holes with same diameter
Flat bottom holes of different depth and diameter
SRb image = 50 µm
PTconst
d CNR SRb
Each row has indications with same CNR = Cdepth/
PTconst - constant human perception threshold 10 d – diameter of just visible hole CNR - contrast to noise ratio SRb – basic spatial resolution (effective pixel size)
Rose approach, 1946
8.3 Radiological Methods
ECNDT Praha, Oct. 2014
Requirements for RT-D Ewert and Zscherpel
8
System Selection for ASTM Standard Practice ASTM Draft, “Practice for the Use of Digital Detector Arrays and Computed Radiology for Aerospace Casting Inspections” requires a minimum basic spatial resolution of the detector (SRbdetector) if images have been taken without magnification or a minimum basic spatial resolution in the magnified image SRbimage).
Effective pixel size of digital image ( SRimage ) b
8.3 Radiological Methods
ECNDT Praha, Oct. 2014
Requirements for RT-D Ewert and Zscherpel
9 9
Verification of CNRmin for ASTM Standard Practice ASTM Draft, “Practice for the Use of Digital Detector Arrays and Computed Radiology for Aerospace Casting Inspections” and E 2698 “Standard Practice for Radiological Examination Using Digital Detector Arrays” require for digital images with DDAs in the 2T IQI hole:
CNRmin 2.5 The ASTM E 2698 value (CNR = 2.5) is optimized for testing of thin objects with thickness < ½” and typical DDAs in the range of 0.1 mm < SRb < 130 µm. Generally, CNRmin for all digital images can be described more accurate by the equation:
CNRmin
10 SRbimage diameterIQI hole
Based on “old” Rose equation as shown before for PTconstant
8.3 Radiological Methods
ECNDT Praha, Oct. 2014
Requirements for RT-D Ewert and Zscherpel
10
Contrast Sensitivity as Required by Different Standards Almost all standards require image quality indicators on each production radiograph. The operator has to evaluate if the required image quality has been achieved. He decides about acceptance or rejection of the production radiographs. Different international standards require different contrast sensitivities, which yields different inspection quality. The thickness sensitivity improves with material thickness. 8.3 Radiological Methods
ECNDT Praha, Oct. 2014
Requirements for RT-D Ewert and Zscherpel
11
Basic Requirements for IQI Visibility of International Standards in Comparison ISO 19232-3
Step Holes, ISO 17636-2, 2013 Plate Holes, ASME BPVC SC V Ar. 2 Plate Holes, ASTM E 1742, E 2104 ASME BPVC SC V, Ar. 2, and ISO 17636-class A are about equivalent ASTM E1742, E 2104, ISO 17636-class B are about equivalent below t=12 mm (1/2“)
ASTM E 1025
ISO 19232-2
8.3 Radiological Methods
ECNDT Praha, Oct. 2014
Requirements for RT-D Ewert and Zscherpel
12
Additional Requirements for Unsharpness and Basic Spatial Resolution In all film radiography standards the permitted geometric unsharpness (Ug) is limited and the film unsharpness is neglected. In digital radiography the detector unsharpness (Udetector) contributes significantly to the image unsharpness (Uim) or total unsharpness (UT). Therefore, the permitted detector unsharpness (Udetector) is limited in different standard systems, which is relevant for detector selection. The detector unsharpness shall be always smaller than the permitted geometric unsharpness (Udetector < Ug). The basic spatial resolution is defined as ½ unsharpness in digital radiography (SRb = ½ U).
8.3 Radiological Methods
ECNDT Praha, Oct. 2014
Requirements for RT-D Ewert and Zscherpel
13
Measurement of Basic Spatial Resolution
8.3 Radiological Methods
ECNDT Praha, Oct. 2014
1m SDD, small focus
• The detector unsharpness udetector shall be controlled by reference exposures with the duplex wire IQI.
Duplex Wire IQI, 90 kV,
Duplex wire IQI ASTM E 2002 Two new ISO 19232-5 wire pairs ASTM E 2002
Requirements for RT-D Ewert and Zscherpel
14
Standard ISO 17636-2 (2013): Table B.13, B.14
8.3 Radiological Methods
ECNDT Praha, Oct. 2014
Requirements for RT-D Ewert and Zscherpel
CR
DDA
Maximum detector or image unsharpness (2SRb)
15
Unsharpness Requirements of Different International Standards – No International Harmonization Unsharpness requirements 0,08
E 1030, 1032
Unsharpness in inch
0,07 0,06
E 1030 (2004)
0,05
MAI E 2104
0,04 0,03
E 1742
0,02
ISO 17636 class A ISO 17636 class B
0,01 0,00 0,0
1,0
2,0
3,0
Thickness in inch
8.3 Radiological Methods
ECNDT Praha, Oct. 2014
4,0
5,0
Co-60, High Energy
Requirements for RT-D Ewert and Zscherpel
6,0
16
Unsharpness Requirements in Different Int. Standards
8.3 Radiological Methods
ECNDT Praha, Oct. 2014
Requirements for RT-D Ewert and Zscherpel
17
EPS= equivalent penetrameter sensitivity (see ASTM E 746, E 747, E 1025)
Qualification of CR-Systems - The EPS Concept EPS – Procedure is proposed accepted for CR qualification in ASTM E 20445/6 draft and the draft on “Practice for the Use of … Computed Radiology for Aerospace Casting Inspections” (USA: MAI – group)
- The EPS (equivalent penetrameter sensitivity) measurement is based on ASTM E 746
A smooth ¾ inch (19 mm) steel plate with a set of plate holes is radiographed at 200 kV in ≥ 1 m distance The exposure is performed with different mAs settings A graph is generated, see next pages The calculation of just visible hole diameter is given by:
d visible
8.3 Radiological Methods
SRbimage PT eff SNR
New formula ECNDT Praha, Oct. 2014
ASTM E 746 PT depends slightly on operator and viewing conditions µeff for 200 kV and 19 mm Fe is about 0,05 mm-1
Requirements for RT-D Ewert and Zscherpel
18
Example: EPS Test with HD CR Scanner at 20 µm Pixel Size
Do you see the holes?
¾” (19 mm) steel plate, 200 kV 1 mm steel step for measurement of µeff
8.3 Radiological Methods
ECNDT Praha, Oct. 2014
CNR 0.8
Requirements for RT-D Ewert and Zscherpel
19
Visibility of EPS holes (E 746)
Do you see the holes?
CNR 0.8 8.3 Radiological Methods
ECNDT Praha, Oct. 2014
Requirements for RT-D Ewert and Zscherpel
20
New Formula for Conversion of SNRN Measurements to EPS Values and Working Range for CR (ASTM Draft E 2033) PT’ is about 2▪100 for visibility of the 2 T hole of IQIs corr. to ASTM E 1025 EPS by ASTM E 746 with 200 kV, t = 19 mm Fe plate and µeff = 0.05 mm-1
EPS method for determination of performance levels Measured EPS,IP 1, Scanner 2
Best fit
Level I
level II
2,6
EPS
2,4 140
EPS
2,2 2
350
Level II: PV 140 – 4095
1,8 1,6
aEPS = 1.16
Level I: PV 350 – 4095
1,4 1,2
90% PVmax
1
0
500
1000
1500
2000
Plateau range 0
23
46
69
2500
3000
3500
4000
137
160
183
450
Pixel value 91
114
206
Dose / mGy EPS= equivalent penetrameter sensitivity (see ASTM E 746, E 747, E 1025)
8.3 Radiological Methods
ECNDT Praha, Oct. 2014
Requirements for RT-D Ewert and Zscherpel
21
PT ' ttestplate
SRbimage µeff SNR
Old CR Classification Scheme System class CEN
Minimum normalised SNR
IP 1/Y
130
IP 2/Y
117
IP 3/Y
78
IP 4/Y
65
IP 5/Y
52
IP 6/Y
43
Note 1: The normalized SNR values of Table 1 are similar to those of EN 584-1. They are calculated by SNR= log(e) (Gradient/Granularity) of Table 1 in EN 584-1. The measured SNR values are calculated from linearised signal data.
8.3 Radiological Methods
ECNDT Praha, Oct. 2014
Classification scheme of ASTM; e 2446, EN 14784-1 and ISO 16371-1 by system classes. The CR systems are classified by the maximum achievable SNRN (IP 1 – 6) and the basic spatial detector resolution SRbdetector (Y).
Requirements for RT-D Ewert and Zscherpel
22
New CR Classification Scheme New classification by performance levels as given in the draft revision of ASTM E 2446 (2014). Additionally, a specified EPS performance is required. CR System Minimum Classification SNRN (normalized to SRb=88.6 µm)
Maximum iSRbdetector value [µm]
Maximum achieved EPS by E 746 [%]
CR Special
200
50
1.00
CR Level I
100
100
1.41
CR Level II
70
160
1.66
CR Level III
50
200
1.92
8.3 Radiological Methods
ECNDT Praha, Oct. 2014
Requirements for RT-D Ewert and Zscherpel
23
Qualification “Spidernet” Graph and Classification Statement
Qualification of CR‐System XY Interpolated SRbdetector for E 2795, ISO 17636-2 users SRbdetector = 145µm 14 12 10 8 6 4 2 0
Achievable SNRN for ISO 17636-2 user aSNRN = 88
Achievable CS for E 2597 user Csa = 0,8
Performance level II
Speed as defined for film ISO Speed@SNRN130 = 400
aEPS@mag 1 = 1,65
Achievable EPS Specific for CR Efficiency@1mGy = 115
Efficiency as defined for DDAs in E 2597
8.3 Radiological Methods
ECNDT Praha, Oct. 2014
Requirements for RT-D Ewert and Zscherpel
24
New Qualification Procedures Required Test for ASTM E 2445/2446
Required Result
Geometric Distortion Laser Jitter Laser Beam Scan Line Integrity Scan column dropout Scanner Slippage Imaging plate Artifacts Erasure Shading or banding
Fail if distortion > 2% Not permitted, Not permitted Not permitted Not permitted Not permitted Fail if > 2% Fail if more than ±10%
Test results shall be reported, classification is possible in case of exceeding the limits PMT Non-linearity Burn-In Spatial Linearity Imaging plate response variation
Result to report
Optional test on request Imaging Plate Fading (no test object required), optional test
Result to report Report fading in %,
8.3 Radiological Methods
ECNDT Praha, Oct. 2014
Requirements for RT-D Ewert and Zscherpel
Report if > 2% Report if > 2% < 2% Report if > ±10%
25
Requirements for Digital Detector Arrays Qualification by ASTM E 2597M
DDA
8.3 Radiological Methods
ECNDT Praha, Oct. 2014
Requirements for RT-D Ewert and Zscherpel
26
BAM 5, 8mm steel FujiFilm IX25 SNRN ~ 265
DDA Technology provides better image quality than film with a special calibration procedure!
Best (slowest) NDT film
Images high pass filtered for better presentation
ECNDT Radiological DDA exposure 8.3 Praha, Oct. 2014 Methods
Requirements for RT-D Ewert and Zscherpel
PerkinElmer 1620 SNRN ~ 1500 Magn. = 3.5
27
Qualification of Digital Detector Arrays - Management of Underperforming Pixels A bad pixel can be corrected if it has at least 5 good neighbors ASTM E 2597M
Definition and Test of Bad Pixels: 6.2.1.1 Dead Pixel 6.2.1.2 Over Responding Pixel 6.2.1.3 Under Responding Pixel 6.2.1.4 Noisy Pixel 6.2.1.5 Non-Uniform Pixel 6.2.1.6 Persistence/Lag Pixel 6.2.1.7 Bad Neighborhood Pixel
6.2.2 Types or Groups of Bad Pixels: 6.2.2.1 Single Bad Pixel 6.2.2.2 Cluster of Bad Pixels 6.2.2.3 A cluster without any CKP is well correctable. 6.2.2.4 A cluster with CKP is labeled a relevant cluster. 6.2.2.5 A single bad line segment is a special irrelevant cluster.
8.3 Radiological Methods
ECNDT Praha, Oct. 2014
Requirements for RT-D Ewert and Zscherpel
28
Qualification of Digital Detector Arrays - Pixel Coverage Recommendations -
ASTM E 2736
8.3 Radiological Methods
ECNDT Praha, Oct. 2014
Requirements for RT-D Ewert and Zscherpel
29
Qualification of Digital Detector Arrays - Management of Underperforming Pixels -
ASTM E 2736
Compensation Principles
8.3 Radiological Methods
ECNDT Praha, Oct. 2014
Requirements for RT-D Ewert and Zscherpel
30
Compensation Principle (II) of ISO 17636-2 Compensation of high detector unsharpness by increased SNR • Unsharp digital systems may be applied for NDT if they enable to compensate the missing sharpness by increased SNR. • That means, achieves a digital system not the required visibility of the separated duplex wires, it can be used for NDT, if one or two single wires more than required (see tables B.1 –B.12 of ISO/DIS 17636-2) can be seen clearly in the digital image for one or two missing duplex wire pairs. Compensation of 3 wires vs. wire pairs requires agreement of contracting parties. • Compensation principle (II): • High detector unsharpness can be compensated by increased SNR
8.3 Radiological Methods
ECNDT Praha, Oct. 2014
Requirements for RT-D Ewert and Zscherpel
31
31
Example : Compensation Principle (II) of ISO 17636-2 Interesting for detectors with higher unsharpness Compensate missing spatial resolution by increased single wire sensitivity: • A lower spatial resolution i.e. a lower double wire score (D) may be compensated by a higher single wire sensitivity i.e. higher single wire score (W). • Max. two (or three) single/double wire scores may be exchanged.
Duplex wire score Single wire score
8.3 Radiological Methods
Not OK D13 W13
ECNDT Praha, Oct. 2014
Required:
OK:
OK
D12 W14
D11 W15
D10 W16
Requirements for RT-D Ewert and Zscherpel
OK by agreement D9 W17 32
Qualification of Digital Detector Arrays - Efficiency Measurement 120 kV, 3 mm Cu
120 kV, 40 mm Al
160 kV
50 kV 90 kV
Qualification by ASTM E 2597M 220 kV, 40 mm Al
8.3 Radiological Methods
ECNDT Praha, Oct. 2014
Requirements for RT-D Ewert and Zscherpel
33
Qualification of Digital Detector Arrays - Image Lag Qualification by ASTM E 2597M
Important for CT and fast measurements as e.g. automated defect recognition (ADR).
8.3 Radiological Methods
ECNDT Praha, Oct. 2014
Requirements for RT-D Ewert and Zscherpel
34
Qualification of Digital Detector Arrays - Contrast Sensitivity by CNR-Step Wedge Measurement Qualification by ASTM E 2597M
Step wedge with notch Measure contrast and noise per step
Contrast sensitivity (CA) at 5% notch in a step wedge by 1/CNR
8.3 Radiological Methods
ECNDT Praha, Oct. 2014
Requirements for RT-D Ewert and Zscherpel
35
Qualification of Digital Detector Arrays - Contrast Sensitivity by CNR-Step Wedge Measurement Qualification by ASTM E 2597M
Working range (4s)
8.3 Radiological Methods
ECNDT Praha, Oct. 2014
Requirements for RT-D Ewert and Zscherpel
36
Qualification of Digital Detector Arrays - SMTR by SNR-Step Wedge Measurement Qualification by ASTM E 2597M
SNR = 250
SNR = 130
Specific Material Thickness Range by SNR limits
Qualification by ASTM E 2597M
8.3 Radiological Methods
ECNDT Praha, Oct. 2014
Requirements for RT-D Ewert and Zscherpel
37
Qualification of Digital Detector Arrays DDAs are qualified by different tests of ASTM E 2597. Five relevant parameters shall be provided:
Basic spatial resolution (SRb) Specific Material Thickness Range Efficiency Image lag Contrast sensitivity
8.3 Radiological Methods
ECNDT Praha, Oct. 2014
Qualification by ASTM E 2597M
Requirements for RT-D Ewert and Zscherpel
3838
Conclusions
Replacement of film radiography by digital techniques in NDT requires careful selection of suitable digital detectors.
Standards on digital radiography were published and revised since 2005 in ISO, CEN, ASTM and ASME, defining requirements for image quality and detector selection.
International standards define different requirements for image quality and detector selection.
The image quality depends on the essential parameters: specific contrast µeff, SNR and basic spatial resolution SRb.
CR is accepted as film replacement technology:
The prove of image quality requires SNRN measurements and IQI visibility (ISO 176362).
CR classification and qualification is under major revision.
DDAs achieve a significantly better contrast sensitivity with suitable detector calibration than film radiography.
Classification will consider SNRN, SRbdetector and EPS
The prove of image quality requires CNR measurements and IQI visibility (ASTM practice E 2798).
DDAs are qualified by different but similar procedures than CR systems, because the DDA detector calibration and image integration influences seriously the qualification, and DDAs provide currently better image quality than CR.
8.3 Radiological Methods
ECNDT Praha, Oct. 2014
Requirements for RT-D Ewert and Zscherpel
39
End
[email protected]
8.3 Radiological Methods
ECNDT Praha, Oct. 2014
Requirements for RT-D Ewert and Zscherpel
40